Aortic valve calcification : in vivo and ex vivo evaluation
Author: Yousry, Mohamed
Date: 2014-09-19
Location: Thoraxaulan, N2:U1, Karolinska University Hospital
Time: 09:00
Department: Inst för molekylär medicin och kirurgi / Dept of Molecular Medicine and Surgery
Abstract
Aortic valve calcification (AVC) or thickening is found in around one fifth of the general population between 65-75 years of age and increasingly thereafter. The process of aortic valve thickening and calcification is not only an aging (wear and tear) process of the valve leaflets. It is now considered to be closely related to atherosclerosis. The presence and the degree of AVC have been shown to have prognostic value in patients with cardiovascular diseases and in the general population. Despite its importance, there is no widely agreed upon scoring system that objectively quantifies AVC.
The aim of this project was to investigate different methods for evaluating AVC, using transthoracic and transoesophageal echocardiography (TTE and TOE), intra-operative assessment of the valve (IOS) and ex vivo evaluation based on computed tomography (CT) of the excised aortic valves and based on valve weight. A 5-grade scoring system was used for the visual assessment of AVC on real-time and still TTE and TOE images of the aortic valve, as well as intra-operatively. Computer-based greyscale measurement (GSM) software was used to obtain a quantitative-ultrasoundbased AVC measure, which was compared with the visual AVC score from TTE and TOE, whereas IOS was used as the gold-standard method. We also aimed at identifying the most suitable ex vivo conditions and CT parameters for optimal AVC scanning by CT in a calcium hydroxyapatite (CaHA) phantom study. The TTE and TOE AVC scores and IOS were compared to the weight and the CT CaHA mineral mass (MM) index of the explanted aortic valves. The study cohorts were recruited among patients undergoing aortic valve replacement because of aortic valve disease and/or ascending aorta aneurysm.
In Study I, which included 185 patients, we showed that the visual evaluation of AVC using real-time TTE images yielded better correlations with IOS than did quantitative still frame measures based on GSM (r = 0.83 vs 0.64, respectively). In Study II, AVC scores based on TTE and TOE realtime images from 169 patients showed strong correlations with IOS (r = 0.83 and 0.82, respectively). GSM-based measures correlated less well with IOS, even for TOE (r = 0.52). In Study II, we also showed that TOE was more accurate than TTE in diagnosing the aortic valve phenotype. In the CTphantom-based, methodological Study III, we identified optimal CT scanning and reconstruction parameters, as well as the most suitable medium (normal saline) for ex vivo tissue CT-based calcium scoring. In Study IV, 155 operatively explanted aortic valves were weighed and scanned ex vivo by CT, and CaHA MM was measured. The CaHA MM exhibited a strong correlation with valve weight (r = 0.91), whereas AVC scores based on TTE, TOE and IOS showed weaker correlations. Conversely, echocardiographic and intra-operative AVC evaluation showed a better correlation with haemodynamic parameters compared with ex vivo CT AVC scoring.
In conclusion, real-time echocardiographic images are crucial for accurate AVC scoring, regardless of whether TTE or TOE is used. Echocardiographic AVC scoring was as accurate as intraoperative assessment, according to valve weight and ex vivo CT. For ex vivo calcium scoring by CT, in addition to scanning and reconstruction parameters, using saline as the surrounding medium seems to be important. TOE is more accurate than TOE in detecting the bicuspid aortic valve.
The aim of this project was to investigate different methods for evaluating AVC, using transthoracic and transoesophageal echocardiography (TTE and TOE), intra-operative assessment of the valve (IOS) and ex vivo evaluation based on computed tomography (CT) of the excised aortic valves and based on valve weight. A 5-grade scoring system was used for the visual assessment of AVC on real-time and still TTE and TOE images of the aortic valve, as well as intra-operatively. Computer-based greyscale measurement (GSM) software was used to obtain a quantitative-ultrasoundbased AVC measure, which was compared with the visual AVC score from TTE and TOE, whereas IOS was used as the gold-standard method. We also aimed at identifying the most suitable ex vivo conditions and CT parameters for optimal AVC scanning by CT in a calcium hydroxyapatite (CaHA) phantom study. The TTE and TOE AVC scores and IOS were compared to the weight and the CT CaHA mineral mass (MM) index of the explanted aortic valves. The study cohorts were recruited among patients undergoing aortic valve replacement because of aortic valve disease and/or ascending aorta aneurysm.
In Study I, which included 185 patients, we showed that the visual evaluation of AVC using real-time TTE images yielded better correlations with IOS than did quantitative still frame measures based on GSM (r = 0.83 vs 0.64, respectively). In Study II, AVC scores based on TTE and TOE realtime images from 169 patients showed strong correlations with IOS (r = 0.83 and 0.82, respectively). GSM-based measures correlated less well with IOS, even for TOE (r = 0.52). In Study II, we also showed that TOE was more accurate than TTE in diagnosing the aortic valve phenotype. In the CTphantom-based, methodological Study III, we identified optimal CT scanning and reconstruction parameters, as well as the most suitable medium (normal saline) for ex vivo tissue CT-based calcium scoring. In Study IV, 155 operatively explanted aortic valves were weighed and scanned ex vivo by CT, and CaHA MM was measured. The CaHA MM exhibited a strong correlation with valve weight (r = 0.91), whereas AVC scores based on TTE, TOE and IOS showed weaker correlations. Conversely, echocardiographic and intra-operative AVC evaluation showed a better correlation with haemodynamic parameters compared with ex vivo CT AVC scoring.
In conclusion, real-time echocardiographic images are crucial for accurate AVC scoring, regardless of whether TTE or TOE is used. Echocardiographic AVC scoring was as accurate as intraoperative assessment, according to valve weight and ex vivo CT. For ex vivo calcium scoring by CT, in addition to scanning and reconstruction parameters, using saline as the surrounding medium seems to be important. TOE is more accurate than TOE in detecting the bicuspid aortic valve.
List of papers:
I. Real time imaging required for optimal echocardiographic assessment of aortic valve calcification. Mohamed Yousry, Anette Rickenlund, Johan Petrini, Tomas Gustavsson, Ulrica Prahl, Jan Liska, Per Eriksson, Anders Franco-Cereceda, Maria J Eriksson, Kenneth Caidahl Clin Physiol Funct Imaging. 2012 Nov;32(6):470-5
Fulltext (DOI)
Pubmed
View record in Web of Science®
II. Aortic valve type and calcification as assessed by transthoracic and transesophageal echocardiography. Mohamed Yousry, Anette Rickenlund, Johan Petrini, Jonas Jenner, Per Eriksson, Anders Franco-Cereceda, Jan Liska, Maria J Eriksson, Kenneth Caidahl Clin Physiol Funct Imaging. 2014 May 29. [Epub ahead of print]
Fulltext (DOI)
Pubmed
III. Quantification of calcium content in small objects by computed tomography: methodological aspects. Mohamed Yousry, Maria J Eriksson, Kenneth Caidahl, Sven Nyrén [Submitted manuscript]
IV. Estimation of aortic valve calcification by echocardiography, surgical scoring, ex vivo computed tomography and valve weight. Mohamed Yousry, Maria J Eriksson, Dianna Bone, Sven Nyrén, Per Eriksson, Anders Franco-Cereceda, Kenneth Caidahl [Manuscript]
I. Real time imaging required for optimal echocardiographic assessment of aortic valve calcification. Mohamed Yousry, Anette Rickenlund, Johan Petrini, Tomas Gustavsson, Ulrica Prahl, Jan Liska, Per Eriksson, Anders Franco-Cereceda, Maria J Eriksson, Kenneth Caidahl Clin Physiol Funct Imaging. 2012 Nov;32(6):470-5
Fulltext (DOI)
Pubmed
View record in Web of Science®
II. Aortic valve type and calcification as assessed by transthoracic and transesophageal echocardiography. Mohamed Yousry, Anette Rickenlund, Johan Petrini, Jonas Jenner, Per Eriksson, Anders Franco-Cereceda, Jan Liska, Maria J Eriksson, Kenneth Caidahl Clin Physiol Funct Imaging. 2014 May 29. [Epub ahead of print]
Fulltext (DOI)
Pubmed
III. Quantification of calcium content in small objects by computed tomography: methodological aspects. Mohamed Yousry, Maria J Eriksson, Kenneth Caidahl, Sven Nyrén [Submitted manuscript]
IV. Estimation of aortic valve calcification by echocardiography, surgical scoring, ex vivo computed tomography and valve weight. Mohamed Yousry, Maria J Eriksson, Dianna Bone, Sven Nyrén, Per Eriksson, Anders Franco-Cereceda, Kenneth Caidahl [Manuscript]
Institution: Karolinska Institutet
Supervisor: Caidahl, Kenneth
Issue date: 2014-08-29
Rights:
Publication year: 2014
ISBN: 978-91-637-6159-1
Statistics
Total Visits
Views | |
---|---|
Aortic ...(legacy) | 836 |
Aortic ... | 280 |
Total Visits Per Month
October 2023 | November 2023 | December 2023 | January 2024 | February 2024 | March 2024 | April 2024 | |
---|---|---|---|---|---|---|---|
Aortic ... | 0 | 0 | 0 | 1 | 0 | 1 | 1 |
File Visits
Views | |
---|---|
Thesis-Mohamed-Yousry.pdf(legacy) | 865 |
Thesis-Mohamed-Yousry.pdf | 463 |
Spikblad_Mohamed_Yousry.pdf(legacy) | 182 |
Spikblad_Mohamed_Yousry.pdf | 21 |
Thesis_Mohamed_Yousry.pdf(legacy) | 20 |
Spikblad_Mohamed_Yousry.pdf(legacy) | 18 |
Spikblad_Mohamed_Yousry.pdf(legacy) | 1 |
Top country views
Views | |
---|---|
United States | 277 |
Sweden | 151 |
Germany | 114 |
China | 112 |
Russia | 14 |
Poland | 12 |
Spain | 11 |
France | 9 |
Ireland | 9 |
Turkey | 9 |
Top cities views
Views | |
---|---|
Sunnyvale | 58 |
Ashburn | 51 |
Stockholm | 35 |
Kiez | 34 |
Beijing | 29 |
Shenzhen | 29 |
Dublin | 8 |
Cracow | 6 |
Des Moines | 6 |
Houston | 6 |